EQUIPMENT
- Perkin-Elmer Model 4100ZL Atomic Absorption Spectrophotometry
- Perkin-Elmer AS-71 Autosampler
- Thallium electrodes discharge lamp
- Perkin-Elmer System 2 EDL power supply
- Hamilton Digital Diluter
- Pyrolitically coated, platform graphite tubes
- 2 mL sample cups
INSTRUMENT PARAMETERS
1 | Wavelength | 276.8 nm |
2 | Slit Width | 0.7 nm |
3 | Slit Height | LOW |
4 | Signal Type | Zeeman AA |
5 | Signal Measurement | Peak Height |
6 | Read Time | 5 seconds |
7 | Read Delay | 0 seconds |
8 | BOC Time | 2 seconds |
9 | Sample Volume | 20 micro L |
10 | Modifier Volume | 10 micro L |
FURNACE PARAMETERS
STEP | TEMP (°C) | Time (sec) | INT. FLOW | |
---|---|---|---|---|
RAMP | HOLD | |||
1 | 110 | 10 | 25 | 250 |
2 | 130 | 10 | 35 | 250 |
3 | 700 | 5 | 30 | 250 |
4 | 1600 | 0 | 5 | 0 |
5 | 2400 | 1 | 2 | 250 |
GAS
- Argon 99.99%
STANDARDS
- Stock Solutions:
- 1000 ppm Thallium: Use certified, Nist traceable Plasma Grade Standard
- 100 ppm Thallium: Pipet 10 mL of 1000 ppm thallium standard into a 100 mL volumetric flask. Dilute to volume with deionized water and mix well.
- 1 ppm Thallium: Pipet 1 mL of 100 ppm thallium standard into a 100 mL volumetric flask. Dilute to volume with deionized water and mix well.
- Instrument Calibration Standards
- 20, 60, and 100 ppb Thallium - use the following settings on the Hamilton digital diluter to make dilutions of the 1 ppm thallium standard to produce the 60 and 100 ppb thallium standard. Make the 20 ppb thallium standard by diluting the 100 ppb thallium standard.
Standard Used Diluter Settings Final Concentration
ppbDiluent Sample 100 ppb 32 80 20 1 ppm 47 30 60 1 ppm 36 40 100
- 20, 60, and 100 ppb Thallium - use the following settings on the Hamilton digital diluter to make dilutions of the 1 ppm thallium standard to produce the 60 and 100 ppb thallium standard. Make the 20 ppb thallium standard by diluting the 100 ppb thallium standard.
MATRIX MODIFIER REAGENTS
- Palladium Nitrate: Pd(NO3)2
- Magnesium Nitrate: Mg(NO3).6H2O
MATRIX MODIFIER PREPARATION
- Weigh 0.1083 g palladium nitrate into a 100 mL amber volumetric flask.
- Add 0.0518 g magnesium nitrate to the 100 mL volumetric flask.
- Dissolve and dilute to volume with deionized water and mix well.
AUTOSAMPLER LOADING
- Load the autosampler tray in the following manner:
Positions 1-34 client samples Position 35 matrix modifier Position 36 deionized water Position 37 20 ppb antimony standard Position 38 60 ppb antimony standard Position 39 100 ppb antimony standard Position 40 quality control sample
PROCEDURE
- Use the same procedure as given in the lead graphite furnace method p. 125- 127, with the following changes:
- Step 2 - change to - Install the thallium electrodes lamp. With the "MOD" button on (green light on), set the power output level to 300 milliamps. Allow the lamp to warm up for 45 minutes. The set current will increase as the lamp warms up. DO NOT READJUST.
- Step 6 - change to: Scroll down to and click on "WATER_TL.GEL"
- Step 15 - change to: Scroll down to and click on "TL_H2O.IDW"
QUALITY CONTROL
- If the r value for the calibration curve is less than .990, remake the standards and recalibrate.
- Analyze the quality control sample before and after the analysis of client samples. Values must be within the control limits established by the supplier of this sample.
REFERENCE
- Methods for Chemical Analysis of Water and Wastes, 1983, Environmental Protection Agency, Thallium Method 279.2 (Atomic Absorption, furnace technique).