PREPARATION PROCEDURE
INSTRUMENT CALIBRATION STANDARDS
- Use only clean pipets. When not in use, store in 5% nitric acid solution.
- Before use, rinse pipets 3 times with deionized water.
- Pour approximately 10 ml. of stock solution into a 30 ml. polypropylene beaker. Swirl and discard.
- Pour approximately 30 ml. of stock solution into the 30 ml. polypropylene beaker.
- Rinse the pipet 3 times with small amounts of the stock solution.
- Pipet the designated amount of stock solution into a 1 liter volumetric flask. If the pipet is to be used again, rinse it 3 times with deionized water before proceeding to the next stock solution. If the pipet is not to be used again, rinse it 3 times with deionized water and return it to the 5% nitric acid storage solution.
- Discard any excess stock solution. Rinse the beaker 3 times with deionized water and allow to air dry before reuse.
- Do Not pipet any stock solution directly from its container.
- Do Not pour an excess stock solution back into its container.
- Do Not use any stock solution that is past its expiration date.
- Record the date prepared, analyst initials, stock solution bottle number, and expiration date on the quality control sheets provided.
- Use a felt tip marker and write the date prepared and the analyst initials on the bottles used to store the calibration standards.